Skip to main content

Case studies

Tracer gas and Tri-Color on the same hood, same day

A pharmaceutical R&D lab needed ASHRAE 110 tracer gas results and wanted to see Tri-Color next to it. One visit gave them the as-installed rating, the trouble spot on the hood, and how it behaves below its setpoint.

Client

Pharmaceutical R&D lab (name withheld)

Sector

Pharmaceutical R&D

Location

Northeast U.S.

Services

Diagnostic

Equipment

ASHRAE 110-2016, SF₆ tracer gas, Q200 analyzer, Tri-Color

Result

Rating, trouble spot and turndown behavior from one visit

3

face velocity setpoints compared: 100, 50 and 70 fpm

Same day

tracer gas and Tri-Color on the same hood

0.00 ppm

tracer gas reading at 70 fpm, at the spot where Tri-Color showed escape

The ask

A rating, plus a second look

The lab's lead engineer needed as-installed ASHRAE 110 tracer gas results on two hoods. He was also interested in dynamic containment testing, so he asked us to run Tri-Color next to tracer gas on one of them: an 8 ft bench hood with a combination sash on VAV controls.

What we did

Both methods, three setpoints

We ran the tracer gas test as installed at 100 fpm, with Tri-Color on the same hood. Then the controls contractor turned the hood down to 50 fpm and back up to 70 fpm, each setpoint confirmed at the opening, and we ran both methods again.

The result

More than a pass or fail

The vertical sash was clean. The hood failed on a 0.27 ppm peak at the center horizontal panel opening, a spot Tri-Color had already flagged before tracer gas was run there. The team left with the rating, the trouble spot, and a clear view of the hood below its setpoint.

What the side-by-side found

Same spot, two different answers

  • 100 fpm, vertical sash: both methods clean.
  • 100 fpm, horizontal panels: Tri-Color showed the laser dipping low first. Tracer gas then failed at the center opening.
  • 50 fpm: both methods failed.
  • 70 fpm: Tri-Color showed escape at the left side. Tracer gas at that same spot read 0.00 ppm.

That isn’t a fault in tracer gas. It’s a limit of it: one probe, one point, nobody moving. Tri-Color watches the whole opening.

Tri-Color at 70 fpm showing haze crossing the laser curtain at the left side of the opening
Tri-Color at 70 fpm: escape at the left side
Tracer gas chart, left position at 70 fpm, flat at 0.00 ppm for the full 300 seconds
Tracer gas at the same spot: 0.00 ppm for 300 seconds

Want the number and the picture?

Tell us about the lab. A fume hood expert will be in touch shortly.