The problem
A spec that asked for more than a face velocity number
The lab architect's specification called for as-installed ASHRAE 110 testing on every hood, and then went further. In its words: "In addition to the static visualizations required in ASHRAE 110-2016, additional tests and videos shall be made according to FumeHoodCertified.com's dynamic test extension to ASHRAE 110." Glen Berry wrote dynamic testing into the spec to make sure the lab would perform the way he designed it, in a teaching lab where students, not technicians, work the hoods.
What FHC did
Full ASHRAE 110 AI on each hood, then Tri-Color
Face velocity profile, smoke visualization, SF₆ tracer gas at left, center and right with a mannequin in the breathing zone, periphery scan and sash movement test, to the spec's 4.0 AI 0.10 criterion. Then the Tri-Color challenges: a person at the hood, sash motion, walk-bys, items leaving the hood, with lasers on the sash plane. Every test was recorded to video and filed against the hood's identification number, as the spec required.
The result
Every hood passed
Each hood has a signed ASHRAE 110-2016 AI report with the numbers, the photos and the Tri-Color findings, and a hood identification number. The lab opened on schedule.